Important Information regarding the MagNA Pure Compact Instrument Operator’s Manual
4
Setup,
page 16
Current Version
Additional Information
Selecting a
Location
The MagNA Pure Compact Instrument requires
very little setup. Choose for the MagNA
Pure Compact Instrument a clean, dry, level,
stable surface within 3 m of a compatible
electrical outlet. To ensure proper ventilation,
leave 10 cm of space behind the instrument and
15 cm at each side of the instrument. No space is
needed at the back of the instrument.
To carry the instrument, place your hands under
the base of the instrument. For this purpose, the
instrument base plate provides four recessed
carrier grips. The weight of the MagNA Pure
Compact Instrument is approx. 60 kg, ensure that
enough manpower is available for transportation.
Description of the Instrument,
page 22
Current Version
Additional Information
UV Light
Protection
The door permits a good view of the interior of
the instrument. During the decontamination
cycle, the door prevents escape of UV light from
the interior.
The Front Door is impervious to UV light from
inside the Instrument, in case the decontamina-
tion function is activated, but permits good view
to the inside of the Instrument. However do not
look at the UV source directly.
Locking of the door is controlled by the software.
Movement of the Robotic Arm is only possible
after the Instrument Door is closed and locked.
Description of the Instrument,
page 29
Current Version
Additional Information
Socket
for power
cable
This is the plug-in for the power cable. There
are two power cables included in the instrument
package, one for US and one for German wall
outlets.
Always connect the Instrument to a grounded
wall outlet.
Performing a Purifi cation Run,
page 43
Current Version
Additional Information
For kit specifi c details regarding the handling of
Reagent Cartridges, please see the correspond-
ing pack insert of the kit.
Reagents might be fl ammable. For kit spe-
cifi c details regarding the handling of Reagent
Cartridges, please consult the package insert of
the respective reagent kit for further information.
Cleaning,
page 81
Current Version
Additional Information
The surface of the MagNA Pure Compact
Instrument as well as removable internal parts
should be cleaned on a weekly basis with a lint
free cloth moistened with deonized water. If a
spill of reagents occurs, the instrument should
be cleaned with a 70% ethanol solution or any
acceptable PCR laboratory cleaner. Turn the
instrument power Off and disconnect or unplug
the power cord before cleaning the instrument.
Mixtures of water and ethanol, that contain
70% ethanol, are highly fl ammable.
MagNA Pure Compact Instrument Operator’s Manual - 2
nd
Addendum
27270610 MagNA Pure Compact OP 1.3 - 2nd Addendum-4S.indd 4
27270610 MagNA Pure Compact OP 1.3 - 2nd Addendum-4S.indd 4
17.06.2010 13:50:33
17.06.2010 13:50:33
Summary of Contents for MagNA Pure Compact Instrument
Page 2: ......
Page 6: ......
Page 11: ......
Page 36: ......
Page 39: ...Prologue MagNA Pure Compact Manual book Seite 3 Dienstag 28 Februar 2006 4 16 16...
Page 45: ...General Overview MagNA Pure Compact Manual book Seite 9 Dienstag 28 Februar 2006 4 16 16...
Page 129: ......