MagNA Pure Compact Operator’s Manual - Version 1.3
72
C
C
After each Leakage Test, do one of the following:
Performing of the Leakage Test
Disposables needed:
䉴
Launch the MagNA Pure Compact soft-
ware.
䉴
On the MAIN MENU Screen that
appears, touch the Maintenance button.
䉴
On the MAINTENANCE Screen that
appears, touch the Leakage Test button.
If the Leakage Test shows
that the O-rings...
Then ...
are leaking and need to be changed
䉴
change the O-rings (as described in Section
C.1.2.2).
䉴
perform the Leakage Test again, to make sure the
new O-rings are working correctly.
are O.K.
do nothing now, but repeat the test after another two
months or 50 purification runs.
Cat. No.
Product
Content
03 561 429 001
MagNA Pure LC O-Ring Maintenance Kit
O-rings (12 x 8), Grease (1)
03 753 166 001
MagNA Pure Compact Tip Tray Kit
10 Tip Trays
Sarstedt
Tubes as recommended in chapter C.5.1 Related Pro-
ducts
User Maintenance
Leakage Test
MagNA Pure Compact Manual.book Seite 72 Dienstag, 28. Februar 2006 4:16 16