Maintenance and Trouble Shooting
73
C
C
䉴
Put 8 new Tip Trays (from the Roche Tip
Tray Kit) into the Tip Tray Rack.
䉴
Pipet 1 ml water into each of 8 Sarstedt
tubes, then insert the tubes into row 1
(Sample Tube row) of the Tube Rack.
䉴
Place 8 empty Sarstedt tubes into the Elu-
tion Tube Rack.
䉴
After doing each of the above steps, touch
the correct confirmation button on the
touch-screen:
• "Filled tubes in sample position" confirmation button
• "Empty tubes in elution position" confirmation button
• "Tip trays inserted" confirmation button
䉴
As you touch each confirmation button,
the "?" symbol on the button changes to a
checkmark.
䉴
After you touch the three buttons, the
Start Test button will become active.
Touch it to start the test.
As the instrument performs the Leakage Test
described below, the screen will display the
time remaining.
Note:
To cancel the test at any time, touch
the Cancel button.
User Maintenance
Leakage Test
MagNA Pure Compact Manual.book Seite 73 Dienstag, 28. Februar 2006 4:16 16