264 • Start Up
EAGLE 2 Operator’s Manual
3. The Battery Voltage Screen appears for a few seconds.
4. The Inert Mode Selection Screen will then be displayed.
Use the AIR
▲
YES or RANGE
▼
SHIFT button to display the mode you wish to
enter. If you have not made a selection after 20 seconds, the instrument will begin to
beep and the LEDs will begin to flash. They will continue until you make a selection.
WARNING:
The EAGLE 2 is not a gas monitoring device until you select a mode
and enter it.
5. If you select Normal Mode, the warm up sequence will continue as described in “Start
Up” on page 27 unless Leak Check or Bar Hole Mode is active. If Leak Check or Bar
Hole Mode is active, the Leak Check/Bar Hole Mode Select Screen will appear
immediately after the Normal Mode selection.
Use the AIR
▲
YES or RANGE
▼
SHIFT button to display the mode you wish to
enter.
6. If you select Inert Mode, the warm up sequence will continue as described in “Start
Up” on page 27. The oxygen alarms in the alarm point screens will be the alarm
points for Inert Mode and not for Normal Mode.
BATTERY MIN:
4.3 VOLTS
BATTERY NOW:
5.2 VOLTS
MODE SELECT
UP/DOWN THEN ENTER
NORMAL MODE
> NORMAL MODE
BAR HOLE MODE
LEAK CHECK MODE