EAGLE 2 Operator’s Manual
Start Up • 263
NOTE:
To change the Normal Mode oxygen other alarm points, select Normal Mode
and see “Updating the Alarm Point Settings” on page 106.
10. Press and release POWER ENTER RESET. The alarm point or alarm operation
(oxygen only) will begin to flash.
11. Use AIR
▲
YES and RANGE
▼
SHIFT to adjust the alarm point to the desired
setting. Keep the following in mind:
•
The low alarm cannot be set higher than the high alarm and the high alarm cannot
be set lower than the low alarm.
•
Any alarm setting can be turned off by adjusting it to its lowest setting. The
setting will be displayed as
OFF.
12. If you want to continue with the change, press and release POWER ENTER RESET
to accept the setting.
If you want to exit this screen without saving any change to the alarms, press and
release DISPLAY ADJUST NO until you return to the Change Alarm Point Settings
Screen.
13. When you are done making changes, use RANGE
▼
SHIFT to move the cursor next
to
END
.
14. Press and release POWER ENTER RESET to save the new settings and return to the
Change Alarm Point Settings Screen.
15. Use RANGE
▼
SHIFT to move the cursor next to
END.
16. Press and release POWER ENTER RESET to return to the main menu.
Start Up
1. Press and briefly hold down the POWER ENTER RESET button. Release the button
when you hear a beep.
2. The LCD will show the following screen for about ten seconds.
OXY 0- 40.0 vol%
INERT ALARM
>LO : 5.0 vol%
HI : 10.0 vol%
END
GAS MONITOR
EAGLE 2