114 • Using Setup Mode
EAGLE 2 Operator’s Manual
3. Use AIR
▲
YES or RANGE
▼
SHIFT to display the desired setting.
4. Press and release POWER ENTER RESET to save the setting and return to the main
menu.
Updating the Calibration Interval
This setting defines the amount of time between calibrations. The time can be set in 1 day
increments. The minimum setting is 1 day and the maximum setting is 365 days. The
factory setting is 90 days.
1. From the main menu, place the cursor next to
CAL INTERVAL
.
2. Press and release POWER ENTER RESET. The Set Calibration Interval Screen
appears.
3. Use AIR
▲
YES or RANGE
▼
SHIFT to display the desired setting.
Tip:
Press and hold AIR
▲
YES or RANGE
▼
SHIFT to rapidly scroll
through the settings.
4. Press and release POWER ENTER RESET to save the setting and return to the main
menu.
Updating LC/BH Mode Setting
With
LC/BH MODE SELECT
set to
LC & BH
, the Mode Select Screen appears when
the unit is turned on. You are able to select from Normal Mode, Leak Check Mode, and
Bar Hole Mode.
With
LC/BH MODE SELECT
set to
BAR HOLE ONLY
, the Mode Select Screen
appears when the unit is turned on. You are able to select from Normal Mode and Bar
Hole Mode.
With
LC/BH MODE SELECT
set to
LEAK CHECK ONLY
, the Mode Select Screen
appears when the unit is turned on. You are able to select from Normal Mode and Leak
Check Mode.
With
LC/BH MODE SELECT
set to
OFF
, the Mode Select Screen does not appear when
the unit is turned on and the unit goes into Normal Mode after the start up sequence.
This setting is factory set to
OFF
when a unit is shipped unless the instrument is ordered
for bar hole measurement or leak checking use. See “Appendix K: Using the EAGLE 2 in
Bar Hole Mode” and “Appendix L: Using the EAGLE 2 in Leak Check Mode” for
discussions of Bar Hole Mode and Leak Check Mode, respectively.
1. From the main menu, place the cursor next to
LC/BH MODE SELECT.
SET CALIBRATION
INTERVAL
UP/DOWN THEN ENTER
90 DAYS