03 Series Operator’s Manual
Setup Mode • 60
11. The display will return to the following screen.
Adjusting the Bump Test Limit Check Setting
You can select how the instrument’s warmup sequence will handle a bump
test being due using the BP--CHK. menu item. This menu item will always
appear in Setup Mode even if Bump Test Function and/or Bump Limit
Display
are set to Off in the 03 Series User Setup Program.
1. At the BP--CHK. screen, press and release the POWER MODE button.
The following screen will display and the current setting will be
displayed.
2. Use the AIR button to scroll through the options. CONF is the factory
setting.
Table 6: Bump Test Limit Check Options
Operation Selection
Description
CANT
Instrument cannot be used until a successful bump test is
performed.
CONF (factory setting)
The user must press and release the AIR button to
acknowledge that a bump test is due or press and release
the POWER MODE button to perform a bump test before
continuing with the warmup sequence.
NONE
The instrument screen indicates a bump test is due upon
startup but continues with the warmup sequence.
ALM--P
BP--CHK.
CONF
BP--CHK.