47 • Calibration Mode
03 Series Operator’s Manual
4. Confirm that the regulator on/off knob is turned all the way clockwise
(closed) and screw the calibration gas cylinder onto the regulator.
5. Push the calibration adapter onto the 03 Series’ sensor face as shown
in Figure 10. The calibration adapter secures itself to the instrument by
latching on to two recesses in the instrument’s rear case. The rubber
boot (if installed) does not need to be removed to install the adapter.
Figure 10: Installing the Calibration Adapter
6. Turn the regulator on/off knob counterclockwise to open it. Calibration
gas will begin to flow.
7. Allow the gas to flow for two minutes.
8. Adjust the display gas reading using the AIR button to match the
calibration cylinder’s target gas concentration. See Step 5 on page 32
for instructions to adjust a parameter using the AIR button. If the
direction of adjustment is changed, the arrows at the bottom of the
screen will change direction.
9. Press and release the POWER MODE button.
10. The 03 Series will make the span adjustment (zero adjustment for OX-
03) and will return to the M--CAL. Screen.
POWER
MODE
CO-03
Calibration
Adapter
AIR
M--CAL.