90
Store Operation
10. Start a measurement. When it fi nishes, a confi rmation screen will
be displayed. Select “Store data” and press the MENU/ENTER key
to store the processing results.
The store process takes about 1 second. When it is completed, the
address is incremented by one step. The stored data includes the
following information: date and time when processing was started,
measurement time, frequency weighting, time weighting (dynamic
characteristics), conditions and processing results, overload and un-
der-range information, other information.
The T-L (time-level graph) display screen is not stored.
Important
Any measurement data present in the currently
displayed address will be overwritten. If the
address is shown in red, it contains data. Take
care not to accidentally overwrite data.
Note
When the number of address for storing data is 1000,
it will not be incremented and it fl ashes on the display.
If you change the address with the or key in
this condition, the fl ashing will stop, and data can
be stored in the newly selected address.
Recalling stored data
The procedure for recalling data stored in memory using Manual mode is
described below.
1. Press the MENU/ENTER key to bring up the menu list screen.
2. Use the
keys to select [Recall] and press the MENU/
ENTER key. The fi le selection screen appears.
3. Use
the
keys to select the data you want to recall and press the
MENU/ENTER key.
4. Use
the
keys to select [Confi rm the data] and press the MENU/
ENTER key. The data stored in memory are displayed.