Chapter 4 To Set the Sample System
RIGOL
MSO8000 User Guide
4-11
LA Memory Depth
Press
Acquire
→
More
to view the memory depth of the current digital channel
under
LA Mem Depth
. The LA memory depth will change with the memory depth of
the analog channel, and cannot be set separately. The maximum LA memory depth is
62.5 Mpts.
Anti-Aliasing
At a slower sweep speed, the sample rate is reduced, and a dedicated display
algorithm can be used to minimize the possibility of aliasing.
Press
Acquire
→
Anti-aliasing
to enable or disable the anti-aliasing function. By
default, anti-aliasing is disabled. In this case, it is more likely to generate waveform
aliasing.
Note:
This function should be enabled when the horizontal time base has a large
deviation from the period of the signal under test (e.g. when the horizontal
time base is greater than five-fold the period of the signal under test).
Horizontal Expansion
Horizontal expansion indicates the reference position that the screen waveform is
referenced to when it is horizontally expanded or compressed in adjusting the
Horizontal
SCALE
knob. In YT mode, press
Acquire
→
More
→
Expand
and
then rotate the multifunction knob
to select the desired reference position. Press
down the knob to select it. You can also press
Expand
continuously to select it or
enable the touch screen to tap the desired reference. The horizontal expansion
reference supported by the oscilloscope includes Center, Left, Right, Trigger, and User.
The default is "Center".
⚫
Center: when the horizontal time base is modified, the waveform will be
expanded or compressed horizontally relative to the screen center.
⚫
Left: when the horizontal time base is modified, the waveform will be expanded
or compressed horizontally relative to the leftmost position of the screen.
⚫
Right: when the horizontal time base is modified, the waveform will be
expanded or compressed horizontally relative to the rightmost position of the
screen.
⚫
Trigger: when the horizontal time base is modified, the waveform will be
expanded or compressed horizontally relative to the trigger point.
⚫
User: when the horizontal time base is modified, the waveform displayed will be
expanded or compressed horizontally relative to the user-defined reference
position.
Summary of Contents for MSO8000 Series
Page 1: ...User Guide MSO8000 Series Digital Oscilloscope Mar 2021 RIGOL TECHNOLOGIES CO LTD...
Page 2: ......
Page 24: ......
Page 72: ......
Page 78: ......
Page 148: ......
Page 202: ......
Page 208: ......
Page 214: ......
Page 248: ...RIGOL Chapter 12 Protocol Decoding 12 14 MSO8000 User Guide...
Page 264: ...RIGOL Chapter 12 Protocol Decoding 12 30 MSO8000 User Guide Sync Error...
Page 284: ......
Page 290: ......
Page 296: ......
Page 302: ......
Page 308: ......
Page 360: ......
Page 383: ...Chapter 21 Remote Control RIGOL MSO8000 User Guide 21 3 5 6 3 4...
Page 394: ......