Chapter 6 Operations and Measurements
RIGOL
MSO8000 User Guide
6-13
two binary bits are as shown in Table 6-4.
Table 6-4 Logic "XOR" Operation
A
B
A^B
0
0
0
0
1
1
1
0
1
1
1
0
Press
Math
→
Math1
→
Operator
to select "A^B":
⚫
Press
Operation
continuously to enable or disable the "XOR" operation
function.
⚫
Press
SourceA
and
SourceB
respectively, and rotate the multifunction knob
to select the channel for SourceA and Source B. Then press down the knob
to select the channel. You can also press
SourceA
and
SourceB
continuously
or enable the touch screen to select the channel. The available channels for
SourceA and SourceB are CH1-CH4 and D0-D15.
Note:
Only the channels that have been enabled currently can be selected.
➢
If Source A (or Source B) selects CH1, press
More
→
Thre.CH1
to set the
threshold of Source A (or Source B) in logic operation. For setting methods,
refer to descriptions in
.
➢
If Source A (or Source B) selects CH2, press
More
→
Thre.CH2
to set the
threshold of Source A (or Source B) in logic operation. For setting methods,
refer to descriptions in
.
➢
If Source A (or Source B) selects CH3, press
More
→
Thre.CH3
to set the
threshold of Source A (or Source B) in logic operation. For setting methods,
refer to descriptions in
.
➢
If Source A (or Source B) selects CH4, press
More
→
Thre.CH4
to set the
threshold of Source A (or Source B) in logic operation. For setting methods,
refer to descriptions in
.
⚫
Press
Offset
, and then rotate the multifunction knob
to adjust the vertical
offset of the operation results. You can also directly rotate
Wave Vertical
POSITION
or enable the touch screen to adjust the vertical offset.
⚫
Press
Size
, and then rotate the multifunction knob
to select the waveform
size. Press down the knob to select it. You can also press
Size
continuously or
enable the touch screen to select it. The available sizes include "Small",
"Medium", and "Large". Also, you can rotate
Wave Vertical
SCALE
to
adjust the waveform size.
⚫
Press
More
→
AutoSetting
to adjust the vertical scale and offset of the
operation results to the optimal value according to the current configuration, so
as to better observe.
⚫
Press
More
→
Sensitivity
to set the sensitivity of the digital signal converted
Summary of Contents for MSO8000 Series
Page 1: ...User Guide MSO8000 Series Digital Oscilloscope Mar 2021 RIGOL TECHNOLOGIES CO LTD...
Page 2: ......
Page 24: ......
Page 72: ......
Page 78: ......
Page 148: ......
Page 202: ......
Page 208: ......
Page 214: ......
Page 248: ...RIGOL Chapter 12 Protocol Decoding 12 14 MSO8000 User Guide...
Page 264: ...RIGOL Chapter 12 Protocol Decoding 12 30 MSO8000 User Guide Sync Error...
Page 284: ......
Page 290: ......
Page 296: ......
Page 302: ......
Page 308: ......
Page 360: ......
Page 383: ...Chapter 21 Remote Control RIGOL MSO8000 User Guide 21 3 5 6 3 4...
Page 394: ......