Chapter 2 Front Panel Operation
RIGOL
User’s Guide for DSA800/E Series
2-95
Freq Count
Figure 2-26 Frequency Counter Measurement
1.
State
Turn on or off the frequency counter.
Key Points:
If no active marker currently exists, turning on the frequency counter will
open a Normal marker automatically.
The frequency readout is more accurate when the frequency counter is
enabled.
The frequency counter measures the frequency near the center
frequency in Zero span mode.
2.
Resolution
Set the resolution of the frequency counter manually or automatically. The
available resolutions are 1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz and 100 kHz.
Table 2-59 Resolution of frequency counter
Parameter
Explanation
Default
1 kHz
Range
1 Hz to 100 kHz
Unit
GHz, MHz, kHz, Hz
Knob Step
10 times
Direction Key Step 10 times
Counter Readings
Summary of Contents for DSA815
Page 1: ...RIGOL User s Guide DSA800 E Series Spectrum Analyzer March 2017 RIGOL TECHNOLOGIES INC...
Page 2: ......
Page 14: ...RIGOL User s Guide for DSA800 E Series XII...
Page 52: ......
Page 189: ...Chapter 3 Remote Control RIGOL User s Guide for DSA800 E Series 3 5 5 6...
Page 238: ......