Chapter 2 Front Panel Operation
RIGOL
User’s Guide for DSA800/E Series
2-57
Table 2-30 Start Line for T-Power Measurement
Parameter
Description
Default
0 us
Range
0 us to Stop Line
Unit
ks, s, ms, us, ns, ps
Knob Step
Sweep Time/600, Min = 1us
Direction Key Step
in 1-1.5-2-3-5-7.5 sequence
5.
Stop Line
Set the right margin (in time unit) of T-Power measurement. The data
calculated under this measurement is between the start line and stop line.
You can use the numeric keys, knob or direction keys to modify this
parameter.
Table 2-31 Stop Line for T-Power Measurement
Parameter
Description
Default
37.5 ms
Range
Start Line to Sweep Time
Unit
ks, s, ms, us, ns, ps
Knob Step
Sweep Time/600, Min = 1us
Direction Key Step
in 1-1.5-2-3-5-7.5 sequence