Chapter 4 Troubleshooting&Message
RIGOL
User’s Guide for DSA800/E Series
4-17
-365
Time out error.
Communicating with USB-GPIB interface converter times out. Please
restart the converter.
300
Print paper error.
Fail to print. Please check whether print papers are available and in the
correct position.
301
Print ink error.
Fail to print. Please check the quantity of ink or confirm whether the ink is
installed correctly.
302
Printer hardware error.
Fail to print. A printer hardware error occurred and please check it.
303
Print file type error.
Fail to print owing to wrong print file type.
304
Unknown print error.
310
Updating error with wrong image file.
An error occurred when updating firmware because the format of the
image file is incorrect or the file has already been damaged.
311
Updating error with wrong revision.
An error occurred when updating firmware because of a revision
incompatibility.
312
Updating error with failing to write flash.
An error occurred when updating firmware owing to a failure to write to
the flash.
320
Invalid file path.
The file path received may not exist or an illegal path format was
received.
321
Invalid input.
An illegal string has been received.
322
File name too long.
The file name is too long. The length should not exceed 48 characters.
323
U-Disk failed to install.
The USB storage device can not be correctly installed and please examine
it for possible damage. For any question, please contact
RIGOL
technical
support.
324
File name existed.
The file name received already exists and please re-input.
325
Input empty.
Summary of Contents for DSA815