Amplitude
Measurement Range
Range
f
c
≥ 10 MHz
DANL to +20 dBm
Maximum Input Level
DC voltage
50 V
CW RF power
attenuation = 30 dB
+20 dBm (100 mW)
Max. damage level*
+30 dBm (1 W)
Displayed Average Noise Level (DANL)
DSA815
Frequency
attenuation = 0 dB, RBW = VBW = 100 Hz, sample
detector, trace average ≥ 50, tracking generator off,
20
℃
to 30
℃
, input impendence = 50 Ω
PA
off
100 kHz to 1 MHz <-90 dBm, <-110 dBm (typ.)
1 MHz to 1.5 GHz <-110 dBm + 6 × (f/1 GHz) dB, <-115 dBm (typ.)
PA
on
100 kHz to 1 MHz <-110 dBm, <-130 dBm (typ.)
1 MHz to 1.5 GHz <-130 dBm + 6 × (f/1 GHz) dB, <-135 dBm (typ.)
Note: *When f
c
≥ 10 MHz, input level > +25 dBm and PA is Off, the protection switch will be on.