Chapter 2 Front Panel Operation
RIGOL
User’s Guide for DSA800/E Series
2-59
Table 2-32 Average Number for ACP Measurement
Parameter
Description
Default
10
Range
1 to 1000
Unit
N/A
Knob Step
1
Direction Key Step
1
2.
Avg Mode
Set the average operation mode to “Exp” or “Repeat” and the default is “Exp”.
When “Exp” is selected, the result is the exponential average of the
current N (N is specified in “Avg Num”) measurement results.
When “Repeat” is selected, the result is the arithmetic average of the
current N (N is specified in “Avg Num”) measurement results.
3.
Main CH BW
Set the bandwidth of the main channel and the power of the main channel is
the power integral within this bandwidth. You can use the numeric keys, knob
and direction keys to modify this parameter.
Table 2-33 Main Channel Bandwidth for ACP Measurement
Parameter
Description
Default
2 MHz
Range
33 Hz to 2.5 GHz
Unit
GHz, MHz, kHz, Hz
Knob Step
Main Channel BW/100, the
minimum is 1Hz
Direction Key Step
in 1-1.5-2-3-5-7.5 sequence
4.
Adj CH BW
Set the frequency width of the adjacent channels.
The adjacent channel bandwidth is related to the main channel
bandwidth and the range available is from mian channel bandwidth/20 to
mian channel bandwidth×20.
You can use the numeric keys, knob or direction keys to modify this
parameter.
Summary of Contents for DSA815
Page 1: ...RIGOL User s Guide DSA800 E Series Spectrum Analyzer March 2017 RIGOL TECHNOLOGIES INC...
Page 2: ......
Page 14: ...RIGOL User s Guide for DSA800 E Series XII...
Page 52: ......
Page 189: ...Chapter 3 Remote Control RIGOL User s Guide for DSA800 E Series 3 5 5 6...
Page 238: ......