6
Appendix: Test Record Form
RIGOL DS70000 Series Digital Oscilloscope Performance Verification Test Record
Form
Model: ________________
Tested by: ________________ Test Date: ________________
Impedance Test Record Form: 1 MΩ Input Impedance
Channel
Vertical Scale
Test Result
Limit
Pass/Fail
CH1
100 mV/div
0.99 MΩ to 1.01 MΩ
500 mV/div
0.99 MΩ to 1.01 MΩ
CH2
100 mV/div
0.99 MΩ to 1.01 MΩ
500 mV/div
0.99 MΩ to 1.01 MΩ
CH3
100 mV/div
0.99 MΩ to 1.01 MΩ
500 mV/div
0.99 MΩ to 1.01 MΩ
CH4
100 mV/div
0.99 MΩ to 1.01 MΩ
500 mV/div
0.99 MΩ to 1.01 MΩ
Impedance Test Record Form: 50 Ω Input Impedance
Channel
Vertical Scale
Test Result
Limit
Pass/Fail
CH1
100 mV/div
48.75 Ω to 51.25 Ω
500 mV/div
48.75 Ω to 51.25 Ω
CH2
100 mV/div
48.75 Ω to 51.25 Ω
500 mV/div
48.75 Ω to 51.25 Ω
CH3
100 mV/div
48.75 Ω to 51.25 Ω
500 mV/div
48.75 Ω to 51.25 Ω
CH4
100 mV/div
48.75 Ω to 51.25 Ω
500 mV/div
48.75 Ω to 51.25 Ω
DC Gain Accuracy Test Record Form: 1 MΩ Input Impedance
Channel Vertical Scale
Test Result
Limit
Pass/Fail
Vavg1
Vavg2
Calculation
Result
[1]
CH1
1 mV/div
≤2%
2 mV/div
≤2%
5 mV/div
≤2%
10 mV/div
≤2%
20 mV/div
≤2%
50 mV/div
≤2%
Appendix: Test Record Form
DS70000 Performance Verification
30
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.