5.5.2
Test Connection Diagram
5.5.3
Test Procedures
WARNING
Before connecting, disconnecting, or moving the test hookup, disable the output of the
signal generator to avoid causing the dangerous voltage.
1. Connect the [RF OUTPUT 50] connector of the RF signal generator DSG3000B
(installed with the OCXO-B08 option) to CH1 of the oscilloscope, as shown in the
figure above.
2. Output a Sine with 10 MHz frequency and 1 Vpp amplitude via DSG3000B.
3. Configure the oscilloscope:
a. Press the front-panel key
to enable CH1.
b. Click or tap the channel status label of CH1 at the bottom of the screen. Then
the Vertical menu is displayed. Then click or tap Probe to enter the Probe
setting menu. Set the probe attenuation ratio to "1X".
c. In the Vertical system menu, click or tap 50 Ω under Impedance to set the
input impedance of CH1 to 50 Ω.
d. Set the vertical scale to 200 mV/div.
e. Set the vertical offset to 0 V. Then close the Vertical system menu.
f. Click or tap the horizontal menu at the top of the screen. Then the Horizontal
system menu is displayed. In this menu, you can set the horizontal timebase to
1 ns/div.
Performance Verification Test
DS70000 Performance Verification
26