5.1.2
Test Connection Diagram
Fluke 9500B
CH1
DS70000
5.1.3
Test Procedures
WARNING
Before connecting, disconnecting, or moving the test hookup, disable the output of the
signal generator to avoid causing the dangerous voltage.
Impedance test when the input impedance is 1 MΩ
1. Connect the active head of Fluke 9500B to CH1 of the oscilloscope, as shown in the
figure above.
2. Configure the oscilloscope:
a. To enable the channel, perform any of the following operations:
- Click or tap the channel status label at the bottom of the screen to enable
the channel.
- Press the front-panel key
to enable the channel, and the backlight of
this key and the corresponding channel key is illuminated.
- In the Vertical menu, select the CH1 tab. Select ON for the Display menu to
turn CH1 on.
b. In the Vertical system menu, click or tap 1 MΩ under Impedance to set the
input impedance of CH1 to 1 MΩ.
c. Set the vertical scale of CH1 to 100 mV/div.
Performance Verification Test
DS70000 Performance Verification
12