5.3
DC Offset Accuracy Test
5.3.1
Specification
DC Offset Accuracy
Specification
≤200 mV/div (±0.1 div±2 mV±1.5% of the offset value)
>200 mV/div (±0.1 div±2 mV±1.0% of the offset value)
5.3.2
Test Connection Diagram
Fluke 9500B
CH1
DS70000
5.3.3
Test Procedures
WARNING
Before connecting, disconnecting, or moving the test hookup, disable the output of the
signal generator to avoid causing the dangerous voltage.
1. Connect the active head of Fluke 9500B to CH1 of the oscilloscope, as shown in the
figure above.
2. Set the impedance of Fluke 9500B to 50 Ω.
3. Configure the oscilloscope:
a. Press
on the front panel to restore the oscilloscope to the default
settings.
b. Click or tap the channel status label at the bottom of the screen. The "Vertical"
system menu is displayed. Click or tap to select 50 Ω for Impedance to set the
input impedance of CH1 to 50 Ω.
Performance Verification Test
DS70000 Performance Verification