Command Systems
RIGOL
Programming Guide for DS1000E, DS1000D Series
2-33
3.
:TRIGger:SLOPe:MODE
Command Format:
:TRIGger:SLOPe:MODE
<mode>
:TRIGger:SLOPe:MODE?
Function Explanation:
The commands set and query the slope condition. <mode> could be
+GREaterthan (positive slope greater than), +LESSthan (positive slope less
than), + EQUal (positive slope equals to), -GREaterthan (negative slope greater
than), -LESSthan (negative slope less than) or –EQUal (negative slope equals
to).
Returned Format:
The query rGREATER THAN, +LESS THAN, +EQUAL, -GREATER THAN,
-LESS THAN
或
-EQUAL.
Example:
:TRIG:SLOP:MODE
+GRE Setup the slope condition as +GREaterthan.
:TRIG:SLOPe:MODE?
The query rGREATER THAN.
4.
:TRIGger:SLOPe:WINDow
Command Format:
:TRIGger:SLOPe:WINDow
<count>
:TRIGger:SLOPe:WINDow?
Function Explanation:
The commands set and query the type of trigger level which can be adjusted by
.
When the slope condition is +GREaterthan, +LESSthan or + EQUal, <count>
could be PA (rising edge Level A), PB (rising edge Level B) or PAB (rising edge
Level AB);
When the slope condition is -GREaterthan, -LESSthan or –EQUal, <count> could
be NA (falling edge Level A), NB (falling edge LevelB) or NAB (falling edge
LevelAB).