Command Systems
RIGOL
Programming Guide for DS1000E, DS1000D Series
2-43
9.
:TRIGger:ALTernation:VIDEO:POLarity
Command Format:
:TRIGger:ALTernation:VIDEO:POLarity
{POSitive|NEGative}
:TRIGger:ALTernation:VIDEO:POLarity?
Function Explanation:
The commands set and query the current video polarity.
Returned Format:
The query returns POSITIVE or NEGATIVE.
Example:
:TRIG:ALT:VIDEO:POL
POS Setup the video polarity as POSITIVE.
:TRIG:ALT:VIDEO:POL?
The query returns POSITIVE.
10.
:TRIGger:ALTernation:VIDEO:STANdard
Command Format:
:TRIGger:ALTernation:VIDEO:STANdard
{NTSC|PALSecam}
:TRIGger:ALTernation:VIDEO:STANdard?
Function Explanation:
The commands set and query the current video trigger standard.
Returned Format:
The query returns NTSC or PAL/SECAM.
Example:
:TRIG:ALT:VIDEO:STAN
NTSC Setup the video standard as NTSC.
:TRIG:ALT:VIDEO:STAN?
The query returns NTSC.
11.
:TRIGger:ALTernation:VIDEO:LINE
Command Format:
:TRIGger:ALTernation:VIDEO:LINE
<value>
:TRIGger:ALTernation:VIDEO:LINE?