15
14
View Window
This window is yellow before applying
blood sample. It gradually turns red when
filled with blood.
Hand Bar
Grip here to insert test strip into meter.
Noble Metal Electrodes
Electrochemical sensor
Indication Symbol
Insert strip with indication symbol
up and toward meter.
The
GM250S Blood Glucose Meter can only be used with the
GS250S Blood Glucose Test Strips .
The application of other Test Strips or control solutions can lead to incorrect results.
Rightest
Rightest
®
®
Sample Port (Sample Entry)
Apply a drop of blood or control solution here.
The
test
request
is
only 1.0
μ
L of blood.
The
GS250S
Rightest
Blood Glucose
Test Strip
®
Electrode Contacts
Sensing signal output terminals.
PRECAUTION
®
-
Re-cap the
Rightest
GS250S Blood Glucose Test Strip vial immediately after taking out a test strip.
®
-
Do not reuse
Rightest
GS250S Blood Glucose Test Strip.
®
-
Do not use expired
Rightest
GS250S Blood Glucose Test Strip.
®
-
Record the date when opening a new
Rightest
GS250S Blood Glucose Test Strip vial for the first time.
Discard the vial of test strips after 3 month from opening.
®
-
Store the
Rightest
GS250S Blood Glucose Test Strip in a cool (39-86
℉
or 4-30
℃
) and dry location (< 90%
relative humidity). Do not expose to direct sunlight or heat.
®
-
For detailed information, please refer to the
Rightest
GS250S Blood Glucose Test Strip Package Insert.
®
- If the
Rightest
GM250S Blood Glucose Meter and GS250S Test Strips are exposed to a high temperature
difference, please wait 30 minutes before measurement.
The
GS250S
Rightest
Blood Glucose
Test Strip
®
Summary of Contents for GM250S