Quality Control Test
Performing a Quality Control Test
OPEN
4
5
6
9
1
2
7
Example of
Expiration Date:
Exp. Date 2013-6
The possible reasons your Control Solution results are out of the range :
®
- Your
Rightest
Control Solution has expired or has been opened 3 months ago.
®
- Your
Rightest
Blood Glucose Test Strip has expired.
®
®
- You left the cap of the
Rightest
Blood Glucose Test Strip vial opened or the
Rightest
Control Solution off for a long
time.
- You didn't perform the test procedure correctly.
®
- Malfunction of the
Rightest
Blood Glucose Meter or GS250S Test Strip.
®
If Control Solution results are out of range, your
Rightest
Blood Glucose Monitoring System may not be working
properly. Repeat the Quality Control Test. If your Control Solution results outside the range still exist, do not use the Blood Glucose
Monitoring System to test your blood glucose and contact your Bionime Customer service.
GS250S
GS250S
GM250S
GM250S
NOTE
- If you want to purchase new normal or high level
Control Solutions, please contact your authorized Bionime
representative.
®
Rightest
PRECAUTION
Each time you open a new bottle of
Control Solution is good for 3 months after opening
the bottle, or until the expiration date printed on the label, whichever comes first.
®
Rightest
®
Rightest
Control Solution, write the discard
date on the label.
®
1. Take one
Rightest
Blood Glucose Test Strip from vial and re-cap the vial cap
immediately.
®
2. Insert the
Rightest
Blood Glucose Test Strip with view window, facing up, into test
strip port.
3. While the blood drop symbol flashing, press and hold the main button for over 3 seconds until
the " " symbol appears.
4. You will see blinking "
"
symbol and " " symbol on the screen, prompting you to apply
®
Rightest
Control Solution.
®
5. Shake the bottle of
Rightest
Control Solution well before opening the cap. Then open the
bottle and put the cap on the table.
®
6. Drip a drop of
Rightest
Control Solution on the top of the cap.
®
7. Gently touch sample entry of the strip with the
Rightest
Control Solution on the top of the
cap.
GS250S
GS250S
8. When you hear a beep (if volume is turned on ) please wait for the test result. The screen will
display the countdown number from 5.