Error Messages and Troubleshooting
Battery Error
1. The " " symbol is blinking when the battery power is low. Please change the meter battery as soon as you can. You
can still do the test.
2. The " " and " " symbols are blinking
.
when the battery is too low. Meter can not do the strip test. Please change
the meter battery immediately.
1a
Error Messages and Troubleshooting
1a
1b
3
Temperature Error
In order to get accurate test result, perform a test between 50~104
℉
(10~40
℃
).
1. When the ambient temperature is 32~48
℉
(0~9
℃
)
or 106~122
℉
(41~50
℃
) the " " warning symbol
will be blinking (1a), you still can do the test but the test result is only for reference, because the test
result under these ranges of temperature might not be correct. In the memory screen this value will be
flagged with “ ” symbol (1b). Repeat the test at an area with temperature between operating range
(50~104
℉
or 10~40
℃
).
2. When you move from the area with temperature outside the operating range of test strip to another area
with temperature inside operating range of test strip, please wait for 30 minutes before you do the test.
®
3. When the temperature is below 32
℉
(0
℃
) or over 122
℉
(50
℃
),
Rightest
Blood Glucose
Meter can not do the test and the " " symbol will blink with thermograph symbol displaying on screen
in this condition. Please move the meter to environment with temperature between 50~104
℉
(10~40
℃
)
and repeat the test after 30 minutes.
GM250S
1b