background image

3-IN-1 BLOOD GLUCOSE MONITOR

Step by step

GL 50 evo

EN

Summary of Contents for GL 50 evo

Page 1: ...3 IN 1 BLOOD GLUCOSE MONITOR Step by step GL 50 evo EN...

Page 2: ......

Page 3: ...ing a sample 21 5 3 Taking a blood sample and measuring the glucose level 22 5 4 Reading the result and marking measurements 24 5 5 Post processing and disposal 25 5 6 Evaluating measured blood glucos...

Page 4: ...for self testing or in a clinical environment by trained personnel It enables you to measure your blood glucose quickly and easily store the measured values and display the average of all measured val...

Page 5: ...the specified Customer Service address A B E D C A 3 in 1 monitor lancing device plug in USB and monitor B USB cover C 10 test strips D 10 sterile lancet needles 5 x 28G blue 5 x 33G purple E 1 practi...

Page 6: ...t the measured values acoustically Display the average of the measured blood glucose values from the last 7 14 30 and 90 days Display the average of the marked measured blood glucose values from the l...

Page 7: ...fficient for n tests LOT Batch designation Art Nr Order number Sterilised by radiation lancets Unit of measurement for blood glucose value Warning see accompanying documents Biohazard risk of infectio...

Page 8: ...self testing Do not share the lancing device or lancet needles with others or amongst various patients risk of infection Use a new sterile lancet needle for each blood sample for single use only Gener...

Page 9: ...ults Relative humidity of more than 90 may lead to inaccurate results Note The Beurer GL50 evo measuring system is intended for measuring capillary and venous whole blood Storage and maintenance Warni...

Page 10: ...aste Electrical and Electronic Equipment If you have any questions please contact the local authorities responsible for waste disposal Notes on handling batteries If your skin or eyes come into contac...

Page 11: ...ce 2 Plug in USB 3 Monitor Front Side 1 Display 2 Slot for test strips 1 2 1 Scroll wheel 2 ON OFF switch 3 Tensioning slider 1 2 3 3 2 Lancing device and lancet needles 1 Cap 2 Protective lancet disc...

Page 12: ...ry 1 State of charge 2 Memory space number 3 Blood sugar unit mg dL or mmol L 4 Time 5 Symbol for marking measurements 6 Date 7 Measured value display Note The monitor is supplied with the following b...

Page 13: ...information is followed Warning Use each test strip only once and for one patient only Handling test strips Note Securely close the test strip box immediately after taking out a test strip Do not use...

Page 14: ...d film packaging never use other containers 4 Initial use and basic settings In order to activate the device move the ON OFF switch into the ON position on off The ON OFF switch is used as a keylock f...

Page 15: ...eel until the arrow is pointing towards Settings then press the scroll wheel All values Average Settings Back 3 Rotate the scroll wheel until the arrow is pointing towards Date time then press the scr...

Page 16: ...to set the timer minute 6 In order to start the timer rotate the scroll wheel to Start and press the scroll wheel Timer Start Stop 3 Setting the alarm You can set an alarm on the blood glucose monito...

Page 17: ...iefly pressing and holding the scroll wheel 2 Rotate the scroll wheel until the arrow is pointing towards Settings then press the scroll wheel All values Average Settings Back 3 Rotate the scroll whee...

Page 18: ...essing the scroll wheel Beep on Beep off Result beep Back Note The result beep transmits the measurement as an acoustic beep signal Example measured value 103 Start signal Beep beep Number 1 Short bee...

Page 19: ...ng towards Turn display then press the scroll wheel Beep on off Ketone on off Turn display Language 8 Selecting the menu language The following menu languages are available English German French Spani...

Page 20: ...fingertip To make the procedure as painless as possible do not take samples directly from the centre of the fingertip but slightly to either side Warning In the event of suspected hypoglycaemia take...

Page 21: ...ve lancet disc by bringing it into a horizontal position and taking it off Retain the protective disc for the safe disposal of the used lancet needle after taking a blood sample 4 Place the cap onto t...

Page 22: ...fore inserting it in the monitor 1 Prepare the test strip Take a test strip from the box film packaging and immediately close it again Use the test strip within three minutes of removal 2 Insert test...

Page 23: ...from an alternative area Carry out another test with blood from your fingertip in order to confirm the test result If you often fail to notice that you have a low blood glucose level carry out a test...

Page 24: ...e performs the blood glucose measurement The monitor counts down for approx five seconds The measurement is then shown on the display MEMORY MEMORY Read your measured value Check again that you have r...

Page 25: ...ssing the scroll wheel 3 The selected marking is now allocated to the measured value and saved to the memory 5 5 Post processing and disposal 1 Remove test strip Remove the test strip from the device...

Page 26: ...trol solution Seek medical advice if dubious results persist Seek medical attention immediately if your symptoms do not correspond to your measured blood glucose values and you have followed all instr...

Page 27: ...spite the preprandial glucose targets having been reached Postprandial blood glucose measurements should be taken one to two hours after the start of a meal as this is when diabetics values are genera...

Page 28: ...ely For the control solution test please observe the additional notes in the instructions for use for the control solution Important Do not use third party control solution Correct functioning of your...

Page 29: ...te Never apply the control solution straight from the bottle to a test strip Reason the remaining solution in the bottle will be contaminated if the top of the bottle comes into contact with the test...

Page 30: ...ible causes include Test strips were exposed to fresh air for too long The test strip box was not closed completely Repeat the test using a new test strip and or a new box of correctly stored test str...

Page 31: ...s 4 You can cancel the process at any time To do so press and hold the scroll wheel or wait until the device switches itself off automatically after 2 minutes 6 2 Displaying average blood glucose valu...

Page 32: ...on page 15 Measurements cannot be taken while the USB stick is connected to a PC The measurements remain saved on the blood glucose monitor when the USB stick is removed from the PC It is not possibl...

Page 33: ...e device on different persons Do not submerge the device in disinfection solutions or any other fluids and make sure that no fluids can enter the device The cap on the integrated lancing device can be...

Page 34: ...hout restriction Low temper ature Temperature of the measuring environment blood glucose monitor or test strip below the permitted range Repeat the test using a new test strip as soon as the measuring...

Page 35: ...rvices Problem the test does not start after inserting the test strip into the device and applying blood Cause Solution Insufficient blood or test strip not filled correctly Repeat test using a new te...

Page 36: ...ble for self testing System function test Each time device is switched on The serial number is located on the device or in the battery compartment EMC This device complies with the European standard E...

Page 37: ...tead of blood The measurement using control solution must be within the result range This range is printed on each test strip box Chemical composition of the control solution The control solution is a...

Page 38: ...1 1 4 2 135 6 7 5 136 3 7 6 1 4 0 1 1 1 3 351 5 19 5 350 8 19 5 2 8 0 2 0 8 System accuracy The GL50 evo blood glucose monitor in comparison to the YSI Three lots of GL50 evo blood glucose test strip...

Page 39: ...within 15 of the values obtained in the medical laboratory with glucose concentrations of at least 100 mg dL 5 55 mmol L You will find further details and information regarding blood glucose results a...

Page 40: ...opamine 2 2 mg dL 0 14 mmol L 5 0 mg dL 0 28 mmol L 1 0 EDTA 5 0 mg dL 0 17 mmol L 2 0 mg dL 0 11 mmol L 2 4 Ephedrine 40 mg dL 2 42 mmol L 3 9 mg dL 0 22 mmol L 2 4 Galactose 20 mg dL 1 11 mmol L 3 1...

Page 41: ...l L 0 8 Tolazamide 2 5 mg dL 0 08 mmol L 2 3 mg dL 0 13 mmol L 1 8 Triglyceride 800 mg dL 9 37 mmol L 7 50 mg dL 0 42 mmol L 4 0 Uric acid 16 5 mg dL 0 98 mmol L 6 6 mg dL 0 37 mmol L 1 8 Xylose 9 5 m...

Page 42: ...y claim shall only be considered if the buyer can provide Beurer or an authorised Beurer partner with a copy of the invoice purchase receipt and the original product The following are explicitly exclu...

Page 43: ......

Page 44: ...ncetas lancette SteriLance Medical Suzhou Inc No 168 PuTuoShan Road New District Suzhou 215153 China GL 50 evo Beurer GmbH S flinger Stra e 218 89077 Ulm Germany www beurer com www beurer healthguide...