Beurer GL50 evo
9
• A haematocrit value (proportion of red blood cells) between 20% and 60% has no significant influence
on the measurement results.
• A very high or low haematocrit value (proportion of red blood cells) may lead to incorrect
measurements. In the event of a very high haematocrit value (above 60%), the displayed blood
glucose value may be too low; in the event of a very low haematocrit value (below 20%), it may be
too high. Consult your GP if you do not know your haematocrit value.
• Do not use the test strips to measure blood glucose values on newborns.
• Do not use NaF or potassium oxalate anticoagulants to prepare for venous blood samples.
• Do not test any severely ill patients using this device.
• Use fresh whole blood only. Do not use serum or plasma.
• Use capillary blood without squeezing the penetration area. Squeezing the area causes the blood to
be diluted with tissue fluid and this may lead to an incorrect measurement.
• Do not use the test strips above an altitude of 7010 metres.
• Very high levels of humidity may influence the test results. Relative humidity of more than 90% may
lead to inaccurate results.
Note
The Beurer GL50 evo measuring system is intended for measuring capillary and venous whole blood.
Storage and maintenance
Warning
• Store the blood glucose monitor and its accessories out of reach of small children and pets. Small
parts, such as lancet needles, cover caps or test strips may be life-threatening when swallowed. If
swallowed, seek medical attention immediately.
• The test strip box contains desiccant, which may irritate the skin or eyes when inhaled or swallowed.
Keep the box out of the reach of children.
The monitor is made from precision and electronic components. The accuracy of the measured values
and service life of the device depend on its careful handling:
• Protect the device and its accessories from impacts, humidity, dirt, marked temperature fluctuations
and direct sunlight. Do not store the device, test strips and control solution in your vehicle, in the
bathroom or in a cooling appliance.
• Do not drop the device.
Repairs
Note
• Do not open the device. Failure to comply with this instruction will void the warranty.
• Do not repair the device yourself. Proper operation can no longer be guaranteed in this case.
• Please contact Customer Services for repairs.