36
Beurer GL50 evo
9 t
echnical
specifications
Dimensions (L x W x H)
123 x 29 x 16 mm
Weight
36 g
Power supply
Li-ion rechargeable battery, 160 mAh
Battery life
150 measurements with fully charged rechargeable
battery
Measured value memory
480 measured values with date/time
Data retained when batteries are changed
Average values
For 7, 14, 30, 90 days
Automatic switch-off
2 minutes after last actuation
Storage/transport temperature
Tempe2°C – +30°C
Relative humidity: < 90%
Operating ranges
Tempe10°C – +40°C
Relative humidity: < 90% non-condensing
Measurement range, glucose
Glucose: 20 – 630 mg/dL (1.1 - 35.0 mmol/L)
Blood sample
Capillary whole blood, venous whole blood
Required volume of blood
0.6 microlitres
Blood glucose measurement
duration
Approx. 5 seconds
Calibration
Plasma
Testing method
Amperometric bio sensor
Usage
Suitable for self-testing
System function test
Each time device is switched on
The serial number is located on the device or in the battery compartment.
EMC
This device complies with the European standard EN 61326 and is subject to specific precautions
with regard to electromagnetic compatibility. Please note that portable and mobile HF communication
systems may interfere with this device. For more details, please contact our Customer Services at
the address indicated.
Test strip functionality
Test strips enable a quantitative measurement of the glucose level in fresh whole blood (capillary or
venous). When the gap for taking blood comes into contact with a drop of blood, it is automatically
filled by simple capillary action. The blood is sucked into the absorbing gap on the test strip and the
monitor measures the glucose level in the blood.
The test is based on the measurement of an electric current that is generated by the chemical reaction
of the glucose with the enzyme glucose dehydrogenase (Aspergillus oryzae) on the strip.