Page
44
Version 02 (2020-04)
ID No. 9230-020-en
5.5.2.1
Terminal strip collection assignment
The connection assignment of the terminal strip is described in
the table below:
Cable
Terminal No.:
Assignment
Strand
Sen
so
r c
ab
le
1
Magnetic field
White
2
Magnetic field sensor -
Brown
3
Shield
Shield and gray wire (via
common insulated wire end
ferrules)
4
Temperature sensor
Blue
5
Black
P
o
we
r
sup
p
ly
Outp
u
t
sig
n
al
6
L+
provided by customer
7
L
−
-O (4-20 mA / HART)
8
+O (4-20 mA / HART)
9
Shield
Shield
NOTICE
The terminals 3 and 9 are connected internally.