SCT-20 Series
26
Visit our website
2.12 Setpoints Programming
1. Press
to enter setpoints and hysteresis settings.
2. Press or until desired setpoint or hysteresis parameter is displayed. Press
.
3. Press or until desired value is displayed. Press
.
4. Press to exit setpoints and hysteresis settings.
These values are set to zero if the calibration is changed significantly. See
2.13 Reserved For The Installer
2.13.1 Menu Locking
This procedure, makes it possible to block the access to any menu on the instrument.
1. Press and hold
then press .
is displayed.
2. Press or until menu to be locked is displayed.
3. Press , and simultaneously for 3 seconds, the display shows
.
The decimal point between the first
and second letter indicates that the menu is now locked. If the operator attempts to enter this menu, access is denied
and the display reads
.
2.13.2 Menu Unlocking
1. Press and hold
then press .
is displayed.
2. Press or until menu to be unlocked is displayed.
3. Press
, and simultaneously for 3 seconds, the display shows
. The decimal point between the first
and second letter is gone, indicating the menu is now unlocked.
2.13.3 Temporary Menu Unlocking
1. Press and hold
then press .
is displayed.
2. Press or until menu to be temporarily unlocked is displayed.
3. Press and simultaneously for 3 seconds, the display shows
. The decimal point between the first and
second letter is gone, indicating the menu is now unlocked and it is possible to enter and modify all menus. By
returning to the weight display, the menu lock is restored.
Parameter
Options
Description
Full Scale
Setpoint; relay switching occurs when the weight exceed the
value set in this parameter. The type of switching is settable.
-Full Scale
Hysteresis, value to be subtracted from the setpoint to obtain
contact switching for decreasing weight. For example with a
setpoint at 100 and hysteresis at 10, the switching occurs at 90
for decreasing weight.
Table 2-17. Setpoints Parameters
Note