Operation
© Rice Lake Weighing Systems
●
All Rights Reserved
47
L
= low half of the DOUBLE WORD forming the number
At the time of writing, the setpoints, hysteresis values are saved to the RAM and will be lost upon the next
power-off; to store them permanently to the EEPROM so that they are maintained at power-on, the ‘99’ command
of the Command Register must be sent.
Weight (Span) Calibration Commands (With Test Weights)
The SCT20-MB calibration can be changed via Modbus. To perform this procedure, the system must be unloaded and the
weight value display reset to zero with the command 100 of the Command Register. A load must then be placed on the system
and the correct weight value must be sent to the registers 40037-40038. To save this value, send the control 101 from the
Command Register. If the operation is completed successfully, the two test weight registers are set to zero.
REGISTER
DESCRIPTION
Saving to EEPROM
ACCESS
40001
Firmware version
—
R
40002
Type of instrument
—
R
40003
Year of Production
—
R
40004
Serial Number
—
R
40005
Active program
—
R
40006
COMMAND REGISTER
NO
W
40007
STATUS REGISTER
—
R
40008
GROSS WEIGHT H
—
R
40009
GROSS WEIGHT L
—
R
40010
NET WEIGHT H
—
R
40011
NET WEIGHT L
—
R
40012
PEAK WEIGHT H
—
RR
40013
PEAK WEIGHT L
—
R
40014
Divisions and Units of measure )
—
R
40015
Coefficient H
—
R
40016
Coefficient L
—
R
40017
Setpoint 1 H
Only after command ‘9’’ of the
COMMAND REGISTER
R/W
40018
Setpoint 1 L
R/W
40019
Setpoint 2 H
R/W
40020
Setpoint 2 L
R/W
40021
Setpoint 3 H
R/W
40022
Setpoint 3 L
R/W
40023
Hysteresis 1 H
R/W
40024
Hysteresis 1 L
R/W
40025
Hysteresis 2 H
R/W
—
40026
Hysteresis 2 L
R/W
—
40027
Hysteresis 3 H
R/W
—
40028
Hysteresis 3 L
R/W
—
40029
Inputs
—
R
40030
OutputS
NO
R/W
40037
Test weight for calibration H
Use with command ‘101’ of the
COMMAND REGISTER
R/W
40038
Test weight for calibration L
R/W
40043
Weight value corresponding to ZERO of
the analog output H
Ony after command ‘99’ of the Command
Register. (Analog Models Only)
R/W
40044
Weight value corresponding to ZERO of
the analog output L
R/W
40045
Weight value corresponding to Full Scale
of the analog output H
R/W
40046
Weight value corresponding to Full Scale
of the analog output L
R/W
Table 3-35. Usable Registers (Modbus-RTU)