![Rice Lake iQube 920i Installation Manual Download Page 34](http://html1.mh-extra.com/html/rice-lake/iqube-920i/iqube-920i_installation-manual_1453911034.webp)
Calibration
31
5.0
Calibration
The
iQUBE
can be calibrated using the
920i
front panel or
Revolution III
. The following sections describe the
calibration procedure for each of the calibration methods.
Figure 5-1. CALIBR Submenu
5.1
920i Front Panel Calibration
The CALIBR submenu (under the SCALES menu,
see Figure 5-1) is used to calibrate the
920i
and the
iQUBE
at the same time.
Three calibration methods can be used to trim the load
cells:
•
Theoretical
•
Cal-Match
•
Manual trimming
5.1.1
Theoretical Calibration
To perform a theoretical calibration of the
iQUBE
using the
920i
front panel, press the
Theoretical
Calibration
softkey. The dead load value is captured
and the cells trimmed based on the capacity and
sensitivity values entered for the cells. Scale capacity
is set at the total of the load cell capacities.
5.1.2
Cal-Match Calibration
To perform a Cal-Match calibration of the
iQUBE
using the
920i
front panel, do the following:
1. Place the indicator in setup mode (display
reads
S
cale Configuration
) and remove all
weight from the scale platform.
2. With the SCALES menu highlighted, press
the
down
key to show the CALIBR parameter.
3. From the CALIBR parameter, press
down
again to show a list of configured platforms
and select the platform to be calibrated.
4. Press
down
to go to the test weight parameter
(
WVAL
). By default, the test weight value is set
to the total load cell capacity. Use the numeric
keypad to enter the test weight, then press
enter
.
5. Press the
Cal-Match
softkey to continue
WZERO
WVAL
W
S
PAN
WLIN
REZERO
POINT 4
POINT 5
POINT 1
POINT 2
POINT
3
VI
S
IBLE
GRAD
S
FORMAT
CALIBR
…
…
CALIB
S
ECTION
S
CORNER
S
WZERO
di
s
play
Li
s
t of
platform
s
PLAT
x
Section
s
di
s
play/tweak
Corner
s
di
s
play/tweak
Deadload
adju
s
tment
weight_value
Theoretical
Calibration
S
CALE
S
S
ERIAL
FEATURE
PFORMT
S
ETPT
S
DIG I/O
ALGOUT
VER
S
S
CALE
x
If
SECTION=PAIRED
mV_value
NOTE: Perform Corner
or Section-Match