Test Mode Operations
29
4.0
Test Mode Operations
Use this mode to test the parameters of the weight indicator.
Note
This procedure should only be performed by a certified technician.
To enter test mode, press the
setup switch
for three seconds. The display will change from
CONFIG to A/DTST
.
T
N
E
R
R
U
C
Y
R
E
T
T
A
B
E
G
A
T
L
O
V
Y
R
E
T
T
A
B
T
S
E
T
D
/
A
uuuuuu
READY
DEFAULT
TRANSMIT U
ECHO R
TEST LCD
IO
$GW6W
W/&G
LR
E$WX
E$W&
G()OW
WQW8
(&+RU
Figure 4-1. Test Mode Operation Menu
Parameter
Description
A/D TEST
Displays A/D Test. Press the
ENTER
key to display the raw count value from A/D converter.
TEST LCD
Press ENTER to light all the segments on the LCD display.
I/O
Display digital input (Remote Switch) and output (Set Point) status.
0 0 0 0 0
OUT2
OUT1
RSW3
RSW2
RSW1
BATTERY
VOLTAGE
Display battery voltage in V.
BATTERY
CURRENT
Display battery charging current in A.
DEFAULT
Default parameters. Press the
ENTER
key to reset configuration and calibration parameters to
factory default values.
NOTE: Not available in panel mode operation
TRANSMIT U
Transmits "U". Press the
ENTER
key to send the U character [ASCII: 85 decimal, 55 hex] to test
the serial line quality and integrity.
Table 4-1. Test Mode Option Parameters