Operating Instructions
21
4. Connect the nasal cannula to the nozzle fitting.
Nozzle fitting is located on the top side of the P2-E6
near the pre-filter.
Connect a nasal cannula to the nozzle fitting on the device.
(pictured)
CAUTION
Ensure that the cannula is routed to prevent it from being pinched or kinked to avoid a
disruption of oxygen flow.
CAUTION
The cannula is designed for disposable use.
5.
Switch on P2-E6 with a short press of the ON/OFF Button.
Press the ON/OFF
button to turn on the concentrator. The device will beep, and
the indicator light will flash.
"Welcome" will appear on the display while the concentrator starts up. The display will
indicate the selected flow setting and power condition. A two-minute warm up time will
initiate. During this period the oxygen concentration is building to the specified value but
may not have yet reached specification. Under special conditions, a longer warm-up time
may be necessary, such as in extremely cold temperatures where the unit was stored or is
being operated.
CAUTION
Oxygen concentration may not reach specification during the two-minute warm up time.
CAUTION
30 seconds after startup, the P2-E6 will enter auto-pulse mode. During these 30 seconds,
inhalation will not work.