14
Alerts-continued
Alarm item
Alarm condition
System process
Display of screen
Battery
Exhausted
Battery cycle > 500
Or health < 50%
Alarm only
Battery Exhausted
Replace
Sieve Bed
Sieve Bed expired or
Sieve Bed chip error
Alarm only
Replace Sieve Bed
Pls Contact provider
Low Input
Voltage
Adapter input
< 17 0v
Switch battery
supply until
adapter input >18
Low Input Voltage
Pls Check Adapter
Absence
Of Breath
No breath detected
continuously > 15S
Alarm only
Absence of Breath
Pls Check cannula
Oxygen
concentration<87
Concentration < 87%
continuously > 300S
Alarm only
Low O2:< 87%
Pls Contact provider
Low Battery
5%
≦
RSOC
≦
!
20%
Without adapter
Alarm only
Low Battery Pls
Charge now
Oxygen
concentration
<50
Concentration < 50%
continuously > 300S
shut down after 30s Low O2:< 50%
Pls Contact provider
Breath Sensor
Fail
Breath Sensor failed
shut down after 30s Breath Sensor Fail
Pls Contact provider
Oxygen Sensor
Fail
Oxygen Sensor
failed
shut down after 30s Oxygen Sensor Fail
Pls Contact provider
Gas Delivery Fail No delivery detected
After injection
shut down after 30s Gas Delivery Fail Pls
Contact provider
Gas Obstruction
Pipe or nasal
blocked
shut down after 30s Gas Obstruction Pls
Contact provider
Tank Pressure
Fail
Tank pressure failed
shut down after 30s Tank Pressure Fail
Pls Check cannula
Sieve Bed Fail
Sieve Bed failure
or invalid
shut down after 10s Sieve Bed Fail
Pls Contact provider
Compressor Fail
Compressor failed
shut down after 10s Compressor Fail
Pls Contact Provider