Serial
Communications
PT12
Manual
7/20/2021
Page
44
of
48
T1.PROBE.ITS90MODE?
T1.PROBE.ITS90MODE=
0
Read/set
the
probe’s
ITS90
mode
0=ITS90
1=ITS90&SR5
2=SR5
Only
T1.PROBE.RTPW?
T1.PROBE.RTPW=
99.99
Read/set
the
probe’s
RTPW
coefficient.
99
<
RTPW
<
101
for
100
Ω
PRT
24
<
RTPW
<
26
for
25
Ω
PRT
T1.PROBE.A?
T1.PROBE.A=
‐
0.001
Read/set
the
probe’s
ITS90
A7
Coefficient
‐
1
≤
A7
≤
1
T1.PROBE.B?
T1.PROBE.B=
‐
0.001
Read/set
the
probe’s
ITS90
B7
Coefficient
‐
1
≤
B7
≤
1
T1.PROBE.C?
T1.PROBE.C=
0.001
Read
the
probe’s
ITS90
C7
Coefficient
‐
1
≤
C7
≤
1
T1.PROBE.A4?
T1.PROBE.A4=
‐
0.001
Read
the
probe’s
ITS90
A4
Coefficient
‐
1
≤
A4
≤
1
T1.PROBE.B4?
T1.PROBE.B4=
‐
0.001
Read
the
probe’s
ITS90
B4
Coefficient
‐
1
≤
B4
≤
1
T1.PROBE.A5?
T1.PROBE.A5=
‐
0.001
Read
the
probe’s
ITS90
A5
Coefficient
‐
1
≤
A5
≤
1
T1.PROBE.B5?
T1.PROBE.B5=
‐
0.001
Read
the
probe’s
ITS90
B5
Coefficient
‐
1
≤
B5
≤
1
Note:
All
of
the
channel
commands
for
T2
‐
T12
are
identical
to
the
commands
above.
Just
replace
T1
with
T2
‐
T12
for
the
desired
channel.