SERENA P GB-R04 del 03-07-17
00/Per Bertani/Manuali d’uso Serena/ Serena con stampante
64/99
10.0.4 REPLACEMENT OF THE VAPORIZER GROUP
(AUTHORIZED TECHNICIAN)
CODE 5177
-
Remove the casing
-
Empty the tanks
-
Remove the two lower springs of the water used tank
-
Disconnect the three tubes of rubber holder and remove the tank, being careful NOT to remove the wires of the
level
-
Unscrew the vaporizer
-
Disconnect the wires and tubes connecting the vaporiser
-
Remove and replace the old with the new vaporizer
-
Replace the wires and tubes in the original position
-
Replace the water tank used taking care to check that inside is perfectly clean
-
Try the sterilizer without the casing (keeping a safe distance) to check any loss
-
Replace the casing and perform a Vacuum Test
10.0.5 REPLACEMENT OF THE ELECTROVALVES GROUP
(AUTHORIZED TECHNICIAN)
CODE
5181
-
Remove the casing
-
Disconnect wires and tubes
-
Remove the old bar and replace the new pre-assembled and complete
-
Replace wires and tubes as originally
-
Check without reassembling the casing the presence of any loss (maintain safe distance)
-
Run the Vacuum Test
Remove the network cables! Only authorized technicians!
10.0.6 REPLACEMENT PUMP LOAD
(AUTHORIZED TECHNICIAN)
CODE 5154
-
Remove the casing
-
Empty the tank of clean water
-
Replace the pump and filter making sure to reconnect all tubes and cables in the original position
-
Check for loss
-
Check the efficiency of conducting
10.0.7 MAINTENANCE OF VACUUM PUMP
(AUTHORIZED TECHNICIAN)
CODE 5153 TWO HEADS
TYPE Sterilizer 23lt.:
-
Remove the casing
-
Unscrew the pump placed under the camera
-
Disconnect tubes and cables
-
Remove the pump and replace the membrane and valves of the three heads
-
Replace the pump with cables and tubes in the original position
-
Perform a vacuum test
TYPE Sterilizer 18lt.:
-
Remove the casing
-
Unscrew the two heads of the pump without removing the pump
-
Replace the valves and the two membranes
-
Replace the casing
-
Perform a vacuum test
FOR BOTH TYPES:
-
By using an indelible pen mark each head
-
Remove the four screws
-
Take the two heads
-
Remove the two centre screws (two plugs are hidden by blacks) and replace the two valves
-
Replace the two heads respecting the sign made with the indelible pen
-
Replace the membranes taking the two ends and unscrew
-
Put the new membranes and screw all the way down
-
Replace the heads
-
Turn the motor fan by hand (full cycle) checking that nothing has hardness
SEE PUMP EXPLODED