SERENA P GB-R04 del 03-07-17
00/Per Bertani/Manuali d’uso Serena/ Serena con stampante
20/99
To clear the message of dirty water you must empty it and fill with water quality.
Proceed as follows:
• Insert the supplied drain hose into the blue speed connection
• Proceed automatically empting the tank of clean water
• Fill the tank with water quality
3.7 AUTOMATIC LOADING
In the SETTINGS choose AUTOMATIC LOADING:
AUTOMATIC
LOADING
99 MIN.
It is recommended to set about 20-30 minutes if the water supply system has no less than 2 atmospheres.
LOADING
000
The valve of the water purifier starts working for the set minutes pressing the START button.
t.o. LOADING
000
T.
o. LOADING means TIME OUT LOADING. It’s a valve opening timer of the purifier with sufficient time for refilling the
tank with clean water to a minimum. The three zeros indicate that no passage of water has been found therefore the
quality has not been checked. In case of normal water passage numbers instead of zeros attest to the quality of the
water.
1. Make sure the tap of the cleaner is opened.
2. The technician must verify the correct operation of the filling device present in the autoclave.
3. For the detailed calibration, refer to the instruction manual of our cleaner.
• Insert the attack provided to the front of the sterilizer purifier (9).
• Connect the plug of the purifier into the plug to the back of the sterilizer
For the first functioning of purifier, check the pressure gauge. The pressure must be at least 2 bars.
Otherwise, extend the filling T.O.
After replacing the cartridges, turn off and then switch on the sterilizer.
Just in case is selected the automatic filling, it is the task manager, to provide the sterilizer a check valve
conforms to IEC61770.