3.4 QUALITATIVE WATER TABLE
FILL THE TANK
CLEAN WATER
PRESS START
010
3.5 CLEAN WATER LOAD
FILLING THE TANK WITH CLEAN WATER
Use only distilled water of high quality! Follow the qualitative water table!
3.6 MANUAL FILLING
Place the tank of distilled water (standard 5 lt.) on the right side of the autoclave at about the level of the car door.
Connect the water hose supplied to the hose connection (14) by inserting the other end into the tank
Press START
In this way the pump is activated and starts filling water
If the water is non-
compliant (> 15μS), the pump stops and the message of dirty water appears on the display.
The filling takes place in two or three times since the command is timed. The pump will stop automatically
when the tank is full and on the display will appear the screen and then the filling.
If you notice that during the filling in the tank the water is running out, the pump stops as you press START.
Provide for a new tank and re-press START to continue and complete the filling.
Disconnect the water supply hose and keep it for the next filling.
DIRTY WATER
EMPTY THE TANK
033
CEN STANDARD DIN EN 285
MAXIMUM VALUE
Evaporation residue
10mg/l
Silicon oxide (SiO
2
)
1mg/l
Iron
0,2mg/l
Cadmium
0,005mg/l
Lead
0,05mg/l
Remains heavy metal (excluded iron cadmium and lead)
0,01mg/l
Chloride
2mg/l
Phosphate
0,5mg/l
Conductivity (at 20° C)
15μs/cm
Ph Value
From 5 to 7
Aspect
colourless, clean and free of sediment
Hardness
0,02 mmol/l
NOTE: the use of feed water containing concentrations higher than those indicated in the table above may significantly reduce the life of the device causing serious
damages to its components and determining the loss of the warranty