iBreeze CPAP Clinical Guide
12
6.6 Clinical Alert Settings
Select Alert menu in the system settings screen to enter the alert settings menu.
Clinical Alert Settings Interface
#
Feature
Description
1
Leakage high
Alert level: Alert1
Setting leakage high alert
Option: OFF / 15S / 30S / 45S / 60S Default: OFF
Alert limit: Total velocity-base velocity>100L/min, trigger time is
the set time.
2
Mask vent holds
blocked
Alert level: Alert1
Setting Mask vent holds blocked alert
Option: ON / OFF Default: OFF
Alert limit: the actual leak flow is less than 10lpm or 30% of the
expected leak flow at present mask pressure and lasts longer
than 60s.
6.7 Clinical Information Interface
Press information key
in clinical standby interface to enter clinical information interface. The
Information Screen provides a summary of the therapy session.
Clinical Information Interface