iBreeze CPAP Clinical Guide
15
7.2 CPAP-V Mode
The CPAP-V mode is a combination of CPAP mode and APAP mode with limited trial days,
which is used to titrate the patient. The device operates in APAP mode with R-Care off in the
preset titration time, during which a P95 pressure is generated, when the titration time expires,
the device operates in CPAP mode, with P95 as suggested prescribed pressure.
CPAP-V Mode Parameters Setting
#
Parameter
Description
1
Pmax
This feature sets upper limit of the treatment pressure in titration session.
The device operates in APAP mode during the titration session, thus
treatment pressure is adapted automatically according to the patient’s
conditions. The initial default setting is 12 cmH2O. You can adjust the
setting from the P min setting to 20 cmH2O. The increment is
0.1/0.5cmH2O.
2
Pmin
This feature sets lower limit of the treatment pressure in titration session.
The device operates in APAP mode during the titration session, thus
treatment pressure is adapted automatically according to the patient’s
conditions. The initial default setting is 6 cmH2O. You can adjust the
setting from 4 cmH2O to the P max setting. The increment is
0.1/0.5cmH2O.
3
Titra time
Set the days that the device operates in APAP mode. When the titra time
expires, the device operates in CPAP mode and P95 is applied as the
prescribed pressure. This screen allows you to adjust the duration of the
Auto-Trial feature in number of days. You can set this from 7 to 28 days.
The default is 7 days.
Note: The CPAP-
V mode can’t be used more than 28 days.