iBreeze CPAP Clinical Guide
10
Icon
Text
Description
P Mask Test
Setting the mask fit testing pressure
Setting Range: 6cmH2O-18, 2cmH2O increment
Default: 10 cmH2O
6.5 Clinical System Settings
Select key
in the clinical standby interface to enter the clinical system settings interface.
Clinical Settings Interface
Icon
Text
Description
Pressure
Unit
Set the pressure unit.
Option: hPa / cmH
2
O Default: cmH
2
O
Screen
Brightness
Adjust the Screen Brightness.
Option: Auto / 1-3 Default: 2
Language
Set the system interface language.
Default: English / Chinese Default: English.
Energy
Saving
Set the Energy Saving function on or off.
When Energy Saving is on,
the screen will be turned off
automatically if there is no operation within 3 minutes in standby
mode, or 30 seconds in therapy mode. If Energy Saving is off, the
screen is always on.
Option: ON / OFF Default: ON
Date
Set the system date.
Note:
1.
Date setting can’t be earlier than the latest time of the report in
the device.
2. The system date is required to reset on the first time start up
when the device is restored the factory default settings.
Date Format
Set the system date format.
Option: YYYY-MM-DD / MM-DD-YYYY / DD-MM-YYYY Default:
YYYY-MM-DD