iBreeze CPAP Clinical Guide
9
Icon
Text
Description
Preheat
Turn Preheat function on or off. When the Preheat is on, the
humidifier starts preheating in the standby mode and the
maximum preheating time is 30 minutes. In the therapy mode,
the preheating stops.
Option: ON / OFF Default: OFF
Note:
1. If the humidifier water level below the limit, Preheat function
will be turned off automatically.
2. If humidifier function is turned off, Preheat function will be
disable.
3
.
Preheat function is not available in the iBreeze 20C.
Auto Ramp
Turn Auto Ramp function on or off.
When Auto Ramp is on, Sleep state can be identified
automatically, sleep begins to start Ramp function, awake do
not start ramp function.
Option: On / OFF Default: OFF
Ramp Time
Set the increase time from Ramp pressure to the setting
therapy pressure.
Option: 0-60 mins, 5 mins increments
Default: 15 mins
Note: if Ramp Time sets to 0 min, Ramp function will be off.
P Ramp
Set the starting pressure of Ramp.
Setting Range: 3 cmH
2
O-Setting pressure, 0.5 cmH
2
O
increments.
Default: 4 cmH
2
O
Auto Start
Turn Auto Start function on or off.
When the Auto Start function is on, the system will start
therapy automatically if a breath with mask is detected. During
therapy, if the mask is removed more than 5 seconds, the
therapy mode will stop.
Option: ON / OFF Default: ON
IPR
Set IPR (Intelligence Pressure Release) level.
Option: OFF / 1-3 Default: 2
Tube Type
Set the Tube type.
Option: 15 mm / 19 mm Default: 15 mm