4.) Press key to make your setting, then press key to select the "LIN COMP"
(linear compensation) function.
Press key into the linear error compensation function, the formula as below:
error value X -(1000/measuring length) = compensation value
(ML) measuring length unit = mm, error unit = μm
Example:
ML=500mm Error = -15μm
-15μ X -(1000/500) = 30μm
the compensation value is 30μm
If set up the linear compensation value of X Axis = 30, through the AXIS key make a
same procedure for Y Axis.
5.) Press key to make you setting, then press key to select the "RAD/DIA"
function.
P r e s s k e y i n t o t h e " R A D / D I A " f u n c t i o n .
- 87 -
R
esson
Technologies
Digital Readout Setup Function
ent
On
Off
RD-11M
R
esson
Technologies
ent
R
I
O
RAD/DIA
On
Off
RD-11M
R
esson
Technologies
ent
R
I
O
SEL AXIS
rAD
rAD
dIA
On
Off
RD-11M
R
esson
Technologies
ent
R
I
O
LIN COMP
On
Off
RD-11M
R
esson
Technologies
ent
R
I
O
RAD/DIA
or
On
Off
RD-11M
R
esson
Technologies
ent
R
I
O
ENTR.PPM
0
0
0
ent
On
Off
RD-11M
R
esson
Technologies
ent
R
I
O
ENTR.PPM
30
0
0
On
Off
RD-11M
R
esson
Technologies
ent
R
I
O
DIRECTN
1
0
0
On
Off
RD-11M
R
esson
Technologies
ent
R
I
O
LIN COMP
ent
ent
ent
ent
Summary of Contents for RD-11M
Page 5: ...1 Basic Functions Resson Technologies On Off RD 11M Resson Technologies ent R I O...
Page 9: ...5 Resson Technologies...
Page 13: ...9 Resson Technologies...
Page 22: ...18 Resson Technologies On Off RD 11M Resson Technologies ent R I O...
Page 27: ...23 Resson Technologies...
Page 32: ...28 Resson Technologies...
Page 40: ...36 Resson Technologies...
Page 46: ...R R R 42 Resson Technologies...
Page 62: ...R R R 58 Resson Technologies...
Page 79: ...75 Resson Technologies...