RAMP
® System Operation
Operator’s Manual, RAMP
®
Reader
35
NOTE:
The RESET LQC TIMER option does not allow the operator to disable the LQC
Timer. In order to turn the LQC Timer off, the steps outlined in Section 2.6.10 should be
followed.
2.6.12 Test Counters
The TEST COUNTERS option maintains a record of the number of Sample, IQC and LQC tests that
the instrument has performed. The counters can be viewed and reset.
From SYSTEM Menu, press [▲] or [▼] until 11. TEST COUNTERS is displayed. Press [Enter] to
select this option. The Reader displays the following prompt (example sample test numbers):
Press [Enter] to display the LQC Counter.
Press [Enter] to display the IQC Counter.
Press [Enter] to display Reset Counters prompt:
Press [Enter] to select NO. If the Counters are to be reset, press [▼] to move the cursor to select the
YES option.
NOTE:
Resetting the Counters will reset
all
the counters (Sample, LQC and IQC) to 0.
2.6.13 Connectivity
Data can be transferred to a computer via the RS232 port on the Reader. Contact Technical Support
on data transfer using HyperTerminal.
2.6.14 Enable Sound
Audible alerts on the Reader can either be disabled or enabled. Press [▲] or [▼] to cycle through the
options. Press [Enter] to save the settings or press [Options] to exit without saving.
RESET >NO
COUNTERS? YES
IQC COUNTER
COUNT NUMBER: 1279
LQC COUNTER
COUNT NUMBER: 341
SAMPLE COUNTER
COUNT NUMBER: 9021