Disinfection
Clean all parts of the mask before disinfecting it as described below.
PASTEURIZATION
Temperature:
70°C
Exposure time:
30 minutes
Result:
High level disinfection
Maximum allowable
cleaning cycles:
10
CIDEX
®
Exposure time:
45 minutes
Rinse:
Per manufacturer’s
instructions
Result:
High level disinfection
Maximum allowable
cleaning cycles:
10
CIDEX O. P. A.
®
Exposure time:
12 minutes
Rinse:
Per manufacturer’s
instructions
Result:
High level disinfection
Maximum allowable
cleaning cycles:
10
STERRAD
®
System 100
Exposure time:
Machine controlled
Result:
Sterilization
Maximum allowable
cleaning cycles:
10
Total is a trademark of Respironics, Inc.
CIDEX, CIDEX O. P. A., and STERRAD System are
registered trademarks of Advanced Sterilization Products.
•
If the swivel does not move freely after disinfection, replace the sealing ring (part no.
452072).
•
Verify function of the flapper valve after processing. If the flapper does not move freely,
replace the mask.
•
Inspect the mask after processing. If any componenets are damaged, replace the Total
Face Mask.
•
If the swivel does not move freely after disinfection, replace the sealing ring (part no.
452072).
•
Verify function of the flapper valve after processing. If the flapper does not move freely,
replace the mask.
•
Inspect the mask after processing. If any componenets are damaged, replace the Total
Face Mask.
•
Slight discoloration of the mask cushion after processing is normal.
•
If the swivel does not move freely after disinfection, replace the sealing ring (part no.
452072).
•
Verify function of the flapper valve after processing. If the flapper does not move freely,
replace the mask.
•
Inspect the mask after processing. If any componenets are damaged, replace the Total
Face Mask.
•
Slight discoloration of the mask cushion after processing is normal.
•
If the swivel does not move freely after disinfection, replace the sealing ring (part no.
452072).
•
Verify function of the flapper valve after processing. If the flapper does not move freely,
replace the mask.
•
Inspect the mask after processing. If any componenets are damaged, replace the Total
Face Mask.
•
A slight odor may be present after processing. This is normal.