Installation Procedure
Page 14
Resonance Technology, Inc.
CinemaVision System Installation / Operation Manual: RTC-CV Rev 19
5.3.5.
Connect the Power Supply cable to the Transducer. Route the cable to one of the round
notches near where the fiber optic bundle exits the housing.
5.3.6.
Carefully pull up the preinstalled protective cable cover film over these connections, making sure
not to pinch any of the cables under the screw posts.
5.3.7.
If the patient monitoring camera option is included, please perform the following steps. If not,
skip to Step 5.2.8.
The MRI/CAM 2020 patient monitoring camera option of the CinemaVision system allows doctors
and technologists to monitor patients undergoing MRI scans. This color camera is MRI-safe and
installed inside the magnet room typically on the wall behind the magnet.
5.3.7.1.
Use the included non-magnetic screws to attach the patient camera to the back wall behind the
magnet.
Put the spare unused
fiber cables back in
the plastic coil.
Rubber grommet
used to secure the
fiber optics bundle
Protective
film