5
Installation Procedure
Page 20
Resonance Technology, Inc.
CinemaVision System Installation / Operation Manual: RTC-CV Rev 19
5.6.
GE Auto Voice Installation
5.6.1.
The GE Auto Voice has an interface box in order to interface the A/V Controller to the MRI signal.
Remove the cable from GE OSG J7 and connect to the Interface box “J7-In”. And connect a cable from
“J7-Out” back to the GE OSG J7 location. The CV Controller receives the signal from interface box via
interlink cable which connects to the controller back panel.
Interface Box
Top View
Interface Box Outputs
5.7.
Siemens Auto Voice Connection
5.7.1.
Remove the (LINE OUT) mini stereo audio cable from the back of the Siemens computer and plug it into
the back of the AV console marked input using a mini stereo extension cable. From the back of the
controller, plug a mini stereo cable (male-male) to output signal to plug back onto the Siemens computer
marked audio out.
Remove Cable from GE OSG J7 and Connect
to Interface Box
Connect from Interface Box to GE OSG J7
Connect Interface Box to back of Controller via Interlink cable
Connect Interface Box to back of Controller via
Interlink cable
Output connection from Interface Box to GE OSG J7