5
Installation Procedure
Page 15
Resonance Technology, Inc.
CinemaVision System Installation / Operation Manual: RTC-CV Rev 19
WARNING: Never use any metal screws that might be attracted by the magnet as they may become
projectiles and cause serious injury or death to anyone inside the magnet room. Remember that the
magnetic field is always present.
5.3.7.2.
Connect the Twin-Ax camera power/signal cable to the camera, and route the cable to the
penetration panel. From the Penetration Panel, Connect the signal to a Transceiver box to run
the camera signal to the Control room to the System Monitor.
WARNING: Avoid creating loops in the cable as it may cause MRI image noise. Also, make
sure to route the cable away from walkways inside the magnet room.
5.3.8.
After making sure the Transducer is powered off. Connect audio headset and video goggle to
transducer. Each latching connector should “click” on both sides when fully seated. If the headset
connection is equipped with a DB-9 connector, make sure the screws on the sides of the connector are
secure.
5.3.9.
Carefully install the cable cover, making sure to align all cables with their respective exit notches in the
housing, and making sure no cables are pinched while installing the cover.
5.4.
Equipment Room Component Interconnection
5.4.1.
Place the Transducer power supply into the location chosen in Step 5.1.4.
This location must be
outside of the 20-Gauss line.
5.4.2.
Connect the round transducer power cable connector to the power supply. Make sure the ring on the
connector is secure.