Installation of the UCC2 and UCC2-2
components
Installation of daughtercards
WARNING:
It is essential that full anti-static precautions are taken before working within the UCC2 controller enclosure.
To install a daughtercard please refer to the daughtercard installation guide (Renishaw part number
Connecting UCC2 cables to and from CMM
Attach all CMM cable connectors to the mating sockets on the UCC2
Secure all cable connectors with the connector jacking screws to ensure physical and electrical connection
Secure all cables for strain relief without reducing access for servicing
Connecting the UCC2 mains supply
This installation must be undertaken by suitably experienced and / or qualified personnel.
The UCC2 contains no user serviceable fuses. It is protected internally from inadvertent current overload by self resetting fuses.
There is no internal fuse on the mains supply of the UCC2. The installer must provide suitable protection. The recommended fuse is a 5 A type
T (anti-surge) to IEC60127/UL248-14, rated to at least 250 V with a high breaking capacity. Its installation must meet the requirements of
IEC61010 and any applicable national wiring regulations for the country of installation.
The UCC2 is isolated from ac power by disconnection of the IEC mains connector on the rear panel. If any additional means of isolation is
required, it must be specified and fitted by the machine manufacturer or the installer of the product. The isolator must comply with IEC61010
and any applicable national wiring regulations for the country of installation.
The use of suitable RCD (residual current device), for automatic disconnection in the event of insulation failure is recommended. This should
be sited within easy reach of the CMM operator and must meet the requirements of IEC61010 and any applicable national wiring regulations
for the country of installation.
UCC2 and UCC2-2 installation guide
www.renishaw.com
Issued 07 2021
12